AS 2472-1996
Superseded
View Superseded by
Valves for medical gas cylinders (Including pin-indexed outlet)
Hardcopy , PDF 1 User , PDF 3 Users , PDF 5 Users , PDF 9 Users
English
01-01-1996
06-30-2017
Specifies non-interchangeable connections for large medical oxygen and nitrous oxide cylinders to comply with AS 2473. For medical oxygen contained in cylinders greater than 5 kg water capacity, Types 10 and 10.5 connections will be used; for all other medical gas cylinders pin-indexed outlet connections are to be used. As the diversity of demands are exhausted for single-pin indexing, double-row pin indexing will be introduced. This Standard introduces pin-indexed connections for all medical nitrous oxide connections, in order to prevent the potentially dangerous connection of industrial carbon dioxide to medical nitrous oxide applications.
| Committee |
ME-002
|
| DocumentType |
Standard
|
| ISBN |
0 7337 0670 3
|
| Pages |
12
|
| ProductNote |
See also AS 2472-1985
|
| PublisherName |
Standards Australia
|
| Status |
Superseded
|
| SupersededBy | |
| Supersedes | |
| UnderRevision |
This standard specified inlet threads, outlet non-interchangeability features, material, testing, and marking requirements for medical gas cylinder valves.NOTE: Although medical gases may also be used for non-medical breathing application, it is not intended that valves specified in this standard be mandatory for non-medical applications.
First published as AS CB4 Supplement 1-1955.
Revised and redesignated AS 2472-1981.
Second edition 1983.
Third edition 1985.
Fourth edition 1996.