IEC 61689:2013

Superseded

Superseded

View Superseded by

Ultrasonics - Physiotherapy systems - Field specifications and methods of measurement in the frequency range 0,5 MHz to 5 MHz

Available format(s)

Hardcopy , PDF

Language(s)

English - French

Published date

02-26-2013

Superseded date

03-08-2022

US$542.00
Excluding Tax where applicable
Summarise this standard

FOREWORD
INTRODUCTION
1 Scope
2 Normative references
3 Terms and definitions
4 List of symbols
5 Ultrasonic field specifications
6 Conditions of measurement and test equipment used
7 Type testing reference procedures and measurements
8 Routine measurement procedure
9 Sampling and uncertainty determination
Annex A (informative) - Guidance for performance
        and safety
Annex B (normative) - Raster scan measurement and
        analysis procedures
Annex C (normative) - Diametrical or line scan measurement
        and analysis procedures
Annex D (informative) - Rationale concerning the beam
        cross-sectional area definition
Annex E (informative) - Factor used to convert the beam
        cross-sectional area (A[BCS]) at the face of the
        treatment head to the effective radiating area (A[ER])
Annex F (informative) - Determining acoustic power through
        radiation force measurements
Annex G (informative) - Validity of low-power measurements
        of the beam cross-sectional area (A[BCS])
Annex H (informative) - Influence of hydrophone effective diameter
Annex I (informative) - Effective radiating area measurement
        using a radiation force balance and absorbing apertures
Annex J (informative) - Guidance on uncertainty determination
Bibliography

IEC 61689:2013 is applicable to ultrasonic equipment designed for physiotherapy containing an ultrasonic transducer generating continuous or quasi-continuous wave ultrasound in the frequency range 0,5 MHz to 5 MHz. This standard only relates to ultrasonic physiotherapy equipment employing a single plane non-focusing circular transducer per treatment head, producing static beams perpendicular to the face of the treatment head. This standard specifies:
- methods of measurement and characterization of the output of ultrasonic physiotherapy equipment based on reference testing methods;
- characteristics to be specified by manufacturers of ultrasonic physiotherapy equipment based on reference testing methods;
- guidelines for safety of the ultrasonic field generated by ultrasonic physiotherapy equipment;
- methods of measurement and characterization of the output of ultrasonic physiotherapy equipment based on routine testing methods;
- and acceptance criteria for aspects of the output of ultrasonic physiotherapy equipment based on routine testing methods. Therapeutic value and methods of use of ultrasonic physiotherapy equipment are not covered by the scope of this standard. This third edition cancels and replaces the second edition published in 2007. It constitutes a technical revision which includes the following significant technical changes with respect to the previous edition:
- restriction introduced of 0,2 W/cm2 effective intensity during hydrophone measurements for treatment heads with ka≤20, to limit the likelihood of cavitation;
- change in the factor Fac, to determine the effective radiating area, from 1,354 to 1,333;
- and change to SI units for terms and definitions.

This publication is to be read in conjunction with IEC 60601-2-5:2009.

Committee
TC 87
DevelopmentNote
Supersedes IEC 60150 (03/2001) To be used in conjunction with IEC 60601-2-5. (08/2007) Stability Date: 2018. (11/2017)
DocumentType
Standard
Pages
124
PublisherName
International Electrotechnical Committee
Status
Superseded
SupersededBy
Supersedes

⚠

A newer version of this standard is available.

View Current Edition View Current Edition