BS EN 60731:2012
Current
The latest, up-to-date edition.
Medical electrical equipment. Dosimeters with ionization chambers as used in radiotherapy
Hardcopy , PDF
English
31-07-2012
INTRODUCTION
1 Scope and object
2 Normative references
3 Terms and definitions
4 General requirements
5 CHAMBER ASSEMBLY performance requirements
6 MEASURING ASSEMBLY performance requirements
7 STABILITY CHECK DEVICE performance requirements
8 Constructional requirements as related to PERFORMANCE
CHARACTERISTICS
9 Marking
10 ACCOMPANYING DOCUMENTS
Annex A (informative) - Values, error and UNCERTAINTY
Annex B (normative) - Test equipment for cable microphony
Annex C (normative) - UNCERTAINTY OF MEASUREMENT
Bibliography
Index of defined terms used in this standard
Annex ZA (normative) - Normative references to international
publications with their corresponding European
publications
Defines the performance requirements of radiotherapy DOSIMETERS, intended for the measurement of ABSORBED DOSE TO WATER or AIR KERMA (and their rates and spatial distributions) in PHOTON, ELECTRON, proton or heavy ion RADIATION FIELDS as USED in RADIOTHERAPY.
| Committee |
CH/62/3
|
| DevelopmentNote |
Supersedes 01/560901 DC and 95/560310 DC. (11/2003) 1997 version along with its amendment remain current & will be withdrawn on 14/03/2015. Supersedes 09/30194820 DC. (07/2012)
|
| DocumentType |
Standard
|
| Pages |
104
|
| PublisherName |
British Standards Institution
|
| Status |
Current
|
| Supersedes |
IEC 60731:2011 specifies the performance requirements of radiotherapy dosimeters, intended for the measurement of absorbed dose to water or air kerma (and their rates and spatial distributions) in photon, electron, proton or heavy ion radiation fields as used in radiotherapy. The dose monitoring systems incorporated in radiotherapy treatment machines are not covered by this standard, neither are the re-entrant ionization chambers used for brachytherapy source calibration and constancy check devices. IEC 60731:2011 is applicable to the following types of dosimeter:
a) field-class dosimeters normally used for
1) in vivo skin surface or intracavitary measurements of dose on patients.
2) in vivo skin surface or intracavitary measurements of dose on patients.
b) reference-class dosimeters normally used for the calibration of field-class dosimeters;
c) scanning-class dosimeters normally used for relative dose distribution measurements with a scanning system such as an automatic water phantom.
The object of this standard is:
- to establish requirements for a satisfactory level of performance for radiotherapy dosimeters;
- to standardize methods for the determination of compliance with this level of performance.
Three levels of performance are specified:
- a lower level of performance applying to field-class dosimeters;
- a higher level of performance applying to reference-class dosimeters;
- a specific level of performance applying to scanning-class dosimeters.
This third edition cancels and replaces the second edition published in 1997 and its Amendment 1 (2002) and constitutes a technical revision. The technical modifications versus the second edition of this standard concerns performance requirements of radiotherapy dosimeters intended for the measurement of absorbed dose to water or air kerma in heavy ion radiation fields and scanning-class dosimeters normally used for relative dose distribution measurements with a scanning system such as an automatic water phantom.
| Standards | Relationship |
| IEC 60731:2011 | Identical |
| NF EN 60731 : 2012 | Identical |
| NBN EN 60731 : 2013 | Identical |
| I.S. EN 60731:2012 | Identical |
| EN 60731:2012 | Identical |
| UNE-EN 60731:2012 | Equivalent |