ASTM F 1456 : 2001
Withdrawn
View Superseded by
Standard Specification for Minimum Performance and Safety Requirements for Capnometers
English
12-31-2010
03-01-2006
CONTAINED IN VOL. 13.01, 2006 Establishes minimum safety and performance requirements for capnometers based on parameters that are achievable within the limits of the existing technology. Applies to capnometers used with adults, children and neonates. Does not apply to devices solely intended for use as cutaneous monitors.
| Committee |
F 29
|
| DocumentType |
Standard
|
| Pages |
10
|
| PublisherName |
American Society for Testing and Materials
|
| Status |
Withdrawn
|
| SupersededBy | |
| Supersedes |
1.1 This clause of the General Standard applies except as follows:
1.1.1 This specification applies to capnometers for use with adults, children, and neonates.
1.1.2 It does not apply to qualitative devices or devices solely intended for use as cutaneous monitors.
1.1.3 It does not apply to devices intended for use in laboratory research applications or nonhuman applications.
1.1.4 This specification does not apply to capnometers intended for use with flammable anesthetic mixtures.
1.1.5 This specification does not apply to devices that are intended for use in single-breath, spot-check, or noncontinuous monitoring applications.
A newer version of this standard is available.