ASTM F 2096 : 2002

Superseded

Superseded

View Superseded by

Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test)

Available format(s)

PDF

Language(s)

English

Published date

04-10-2002

Superseded date

09-30-2020

Superseded by

ASTM F 2096 : 2004

US$83.00
Excluding Tax where applicable

Committee
F 02
DocumentType
Test Method
Pages
5
PublisherName
American Society for Testing and Materials
Status
Superseded
SupersededBy
Supersedes

1.1 This test method covers the detection of gross leaks in medical packaging. Method sensitivity is down to 250 m with an 81 % probability (see Section 11). This test method may be used for tray and pouch packages.

1.2 The sensitivity of this test method has not been evaluated for use with porous materials other than spunbonded polyolefin or with nonporous packaging.

1.3 This test method is destructive in that it requires entry into the package to supply an internal air pressure

1.4 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.

1.5

US$83.00