ASTM F 2063 : 2012

Superseded

Superseded

A superseded Standard is one, which is fully replaced by another Standard, which is a new edition of the same Standard.

View Superseded by

Standard Specification for Wrought Nickel-Titanium Shape Memory Alloys for Medical Devices and Surgical Implants

Available format(s)

PDF

Language(s)

English

Published date

01-12-2012

Superseded date

24-08-2018

Superseded by

ASTM F 2063 : 2018

£57.74
Excluding VAT

Committee
F 04
DocumentType
Standard
Pages
6
PublisherName
American Society for Testing and Materials
Status
Superseded
SupersededBy
Supersedes

1.1 This specification covers the chemical, physical, mechanical, and metallurgical requirements for wrought nickel-titanium bar, flat rolled products, and tubes containing nominally 54.5- to 57.0-weight percent nickel and used for the manufacture of medical devices and surgical implants.

1.2 Requirements are for mill product, measuring 5.50 to 94.0 mm [0.218 to 3.70 in.] diameter or thickness. Mill product is not intended to have the final shape, final surface finish, or final properties of the medical device, implant, or their components. Finished NiTi cold-worked tube should be considered under Specification F2633.

1.3 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.

ASTM E 3098 : 2017 Standard Test Method for Mechanical Uniaxial Pre-strain and Thermal Free Recovery of Shape Memory Alloys
ASTM F 2004 : 2017 Standard Test Method for Transformation Temperature of Nickel-Titanium Alloys by Thermal Analysis
ASTM F 2633 : 2013 Standard Specification for Wrought Seamless Nickel-Titanium Shape Memory Alloy Tube for Medical Devices and Surgical Implants
ASTM E 3097 : 2017 Standard Test Method for Mechanical Uniaxial Constant Force Thermal Cycling of Shape Memory Alloys

£57.74