• Shopping Cart
    There are no items in your cart
We noticed you’re not on the correct regional site. Switch to our AMERICAS site for the best experience.
Dismiss alert
Meet Inform Select. Always-on access to the Standards your organisation relies on. Explore Inform Select

UNE-EN IEC 61326-3-2:2018

Current

Current

The latest, up-to-date edition.

Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 3-2: Immunity requirements for safety-related systems and for equipment intended to perform safety-related functions (functional safety) - Industrial applications with specified electromagnetic environment (Endorsed by Asociación Española de Normalización in January of 2019.)

Available format(s)

Hardcopy , PDF

Language(s)

English

Published date

01-01-2019

£76.62
Excluding VAT

This part of IEC 61326 covers all equipment within the scope of IEC 61326-1, but is limited to

systems and equipment for industrial applications within a specified electromagnetic

environment and intended to perform safety functions as defined in IEC 61508 with SIL 1-3.



The electromagnetic environments encompassed by this product family standard are

industrial, both indoor and outdoor, and based on the requirements of the process industry,

specifically chemical/petrochemical/pharmaceutical manufacturing plants using the mitigation

measures given in Annex C. The difference between the electromagnetic environment

covered by this standard compared to the general industrial environment (see IEC 61326-3-1)

is due to the mitigation measures employed against electromagnetic phenomena leading to a

specified electromagnetic environment with test values that have been proven in practice.



The environment of industrial application with a specified electromagnetic environment

typically includes the following characteristics:

industrial area with limited access;

limited use of mobile transmitters;

dedicated cables for power supply and control, signal or communication lines;

separation between power supply and control, signal or communication cables;

factory building mostly consisting of metal construction;

overvoltage/lightning protection by appropriate measures (for example, metal construction of the building or use of protection devices);

pipe heating systems driven by a.c. main power;

no high-voltage substation close to sensitive areas;

presence of CISPR 11 Group 2 ISM equipment using ISM frequencies only with low power;

competent staff;

periodical maintenance of equipment and systems;

mounting and installation guidelines for equipment and systems.



Equipment and systems considered as proven-in-use according to IEC 61508 or prior use

according to IEC 61511 are excluded from the scope of IEC 61326-3-2.



Fire alarm systems and security alarm systems intended for protection of buildings are

excluded from the scope of IEC 61326-3-2.

Committee
CTN 203/SC 65
DocumentType
Standard
ISBN
978-2-8322-4213-1
Pages
48
ProductNote
THIS STANDARD ALSO REFERS TO NAMUR Recommendation NE 21:2012, NAMUR Recommendation NE 93:2003, NAMUR Recommendation NE 98:2007
PublisherName
Asociación Española de Normalización
Status
Current

Standards Relationship
BS EN IEC 61326-3-2:2018 Equivalent
EN IEC 61326-3-2:2018 Identical
IEC 61326-3-2:2017 Identical

£76.62
Excluding VAT